Body' Analysis Turn the power switch of the instrument to the off position, turn the voltage knob to the minimum value, and select the appropriate voltage and current stabilization method and voltage and current range. The DC output end is connected to the two electrodes of the electrophoresis tank using wires. The polarity must be connected correctly. Use the voltage adjustment button to adjust to the required voltage and set the end time of electrophoresis. When electrophoresis is entering work, it should not come into contact with its electrodes, electrophoresis objects, etc. that are charged. Part of the electrophoresis tank, do not put things in the electrophoresis tank. If you need to pick and place items, you should turn off the power first before operating.
Body' Body' After the instrument is powered on and calibrated, the output wire plug should not be temporarily added or removed. When using multiple slots in association, it should be ensured that the total current does not exceed the rated current of the instrument and cannot be overloaded. If the instrument detects electrophoresis input conditions, the load should be connected before turning on the instrument under steady flow conditions to avoid artificial damage. Also, electrophoresis with different media supports should not be performed on the same electrophoresis instrument at the same time.